Atmospheric phase and phase calibration considerations

Phase calibration at 3-mm wavelength is significantly more difficult than at the longer wavelengths. This is because the atmospheric phase errors are comparatively larger, the secondary calibrator grid is sparser and the system sensitivity is poorer. To use a tolerably nearby calibrator, you will need to use a calibrator that is weaker than you might wish - particularly if you wish to observe in a narrow bandwidth. Alternatively you might use a nearby strong SiO maser rather than continuum point source phase calibrator.

These lead to a different approach to calibration compared to that at longer wavelengths. At 3-mm wavelengths, the dual frequency band capability of the ATCA can be used to good effect in the calibration process. When sensitivity to calibrators is an issue, one of the two bands can be specifically set aside for calibration purposes. The correlator configuration can be selected so that one of the bands is either 128 MHz wide (and so as sensitive as possible), or suitable for observing an SiO maser. This calibration band can also be used for reference pointing during the course of the observation.

When using one of the bands for calibration purposes, the two bands need to be jointly calibrated.

In a typical 3-mm observations, in addition to the target source, your will observe a bandpass calibration, a secondary calibrator and a flux density calibrator.

SiO masers should be used with caution: they are often strongly polarised, and they may be structurally complex. Note, also, that the ATCA limits the dual observing bands to be set within 2.7 GHz of each other. If you use an SiO source as a calibrator, the part of the 3-mm band that is accessible to the observation is limited.

Miriad manager
2016-06-21